CO2 and Temperature Monitor ZG1163
  • CO2 and Temperature Monitor ZG1163

ZG1163 CO2 monitor is smart, compact and easy-to-use. In addition to measuring the CO2 concentration, It can also measure the ambient temperature. This product is developed to detect the presence of CO2 in ambient air and help people to take care of Indoor Air Quality. It can be widely used in the office building, school, exhibition, shopping mall, meeting room, fitness center, restaurant and other pubic place where personal comfort, healthy is important.

Details

There are several Modes which can be adjusted by user. These modes are ALTI Mode, ALARM Mode, OUTSIDE Mode, CALI Mode, DATALOGGER Mode, MaxMin Mode and rcFS Mode in sequence.

Features:
1)The built-in Data logger can store 48 sets of CO2 and temperature in the past 24 hours; one log per 30 minutes.
2)Max/Min mode can record the maximum and minimum concentration of CO2 since the device has been last turned on.
3)The Alarm mode the Alarm Level can be adjustable by user.
4)The RCFS Mode can recover the original factory settings after the CO2 device has been recalibrated, altered, or damaged.

Specification:
Measurement Range:0-3,000 ppm display
Display Resolution:1ppm at 0~1,000ppm; 5ppm at 1,001~2,000ppm; 10ppm at 2,001~3,000ppm
Accuracy:0~2,000ppm: ±70 ppm or ±5% of reading, whichever is greater; over 2000ppm: +/-7%
Temperature Dependence: ±0.2% of reading per °C or ±2 ppm per °C, whichever is greater, referenced to 25°C
Pressure Dependence: 0.13% of reading per mm Hg (Corrected via user input for altitude)
Repeatability: ±20 ppm @ 400ppm
Response Time: <2min for 63% of step change
Warm-Up Time: <60 seconds at 22°C
Zone LED Display: Green:<800ppm; Yellow:800~1200ppm;Red: >1200ppm
Temperature Specification
Temperature Range: 0°C to 50°C (32°F to 122°F) display
Display Resolution: 0.1°C (0.1°F)
Display Options: °C/°F
Accuracy: ±1°C( ±2°F) When the fan blows to the device directly, the accuracy of temperature is + / -1.5 degC
Response Time: 20-30 minutes (case must equalize with environment)
RH Specification
Measurement Range: 20%-90% RH
Display Resolution: 1%RH
Accuracy: ±5%RH@23°C
Response time: <5 min for 63% of step change
Operating Conditions
Operating Temperature: 0°C to 50°C (32°F to 122°F)
Humidity Range: 0 ~ 95% RH non-condensing
Storage Temperature: -20°C to 60°C( -4°F to 140°F)
Power Supply: 100 ~ 240 VAC 50 / 60 Hz,6 VDC from external AC/DC adapter which
is included in package (Use specified AC adapter only)